Numerical Research of the Measurement Error of Temperature Thermocouples with the Isolated Seal / Yu. K. Atroshenko, A. A. Bychkova, P. A. Strizhak
Уровень набора: (RuTPU)RU\TPU\network\4526, MATEC Web of ConferencesЯзык: английский.Резюме или реферат: Mathematical models of heattransfer are developed for an assessment of measurement errors of temperature by thermocouples with an isolated and uninsulated seal. Dependences of necessary time of heating (for authentic measurement) for the thermocouples with an isolated seal manufactured of different materials are set. It is shown that for thermocouples with isolated seal minimum necessary duration of heating up slightly exceeds this index for thermocouples with an uninsulated seal..Примечания о наличии в документе библиографии/указателя: [References: 11 tit.].Тематика: электронный ресурс | труды учёных ТПУ | численные исследования | погрешности измерения | термопары | технологические процессы | печать Ресурсы он-лайн:Щелкните здесь для доступа в онлайн | Щелкните здесь для доступа в онлайнНет реальных экземпляров для этой записи
Title screen
[References: 11 tit.]
Mathematical models of heattransfer are developed for an assessment of measurement errors of temperature by thermocouples with an isolated and uninsulated seal. Dependences of necessary time of heating (for authentic measurement) for the thermocouples with an isolated seal manufactured of different materials are set. It is shown that for thermocouples with isolated seal minimum necessary duration of heating up slightly exceeds this index for thermocouples with an uninsulated seal.