Calculating Parameters of Pinhole in Collimation System of Albedo Computed Tomography of Steel Products / E. E. Zhuravsky, B. I. Kapranov, D. S. Belkin [et al.]
Уровень набора: Russian Journal of Nondestructive TestingЯзык: английский.Страна: .Резюме или реферат: We examine specific features of the collimation system of a tomograph for testing steel products based on Compton backscattering. The device is described and the parameters of the primary pinhole collimator of the collimation system are calculated. The device allows reliable scanning of a steel test object by simply displacing the pinhole perpendicular to the X-ray axis from the tube anode within 50 mm with a focal length of 163 mm..Примечания о наличии в документе библиографии/указателя: [References: 12 tit.].Аудитория: .Тематика: электронный ресурс | труды учёных ТПУ | compton tomography | collimation system | pinhole | nondestructive testing | radiation testing | instrument making | компьютерная томография | коллимационные системы | неразрушающий контроль | радиационные испытания | дефектоскопия Ресурсы он-лайн:Щелкните здесь для доступа в онлайнНет реальных экземпляров для этой записи
Title screen
[References: 12 tit.]
We examine specific features of the collimation system of a tomograph for testing steel products based on Compton backscattering. The device is described and the parameters of the primary pinhole collimator of the collimation system are calculated. The device allows reliable scanning of a steel test object by simply displacing the pinhole perpendicular to the X-ray axis from the tube anode within 50 mm with a focal length of 163 mm.
Для данного заглавия нет комментариев.
Личный кабинет оставить комментарий.